Configuration
6-46 Software Installation & Configuration
Bed Config Bed Config is used for two purposes -- to format Patient Sectors on the Main Screen and to
assign patient monitors to Patient Sectors. The Bed Config window is shown in Figure 6-
14.
Figure 6-14 Bed Config Window
Description Fields in Bed Config are:
Field Description
Fast Alarm Display Fast alarm strip displays in the Patient Window when the Silence button has been clicked after an
alarm has occurred.
• Disabled means that a fast alarm strip will not display.
• Red and Yellow means that a fast alarm strip will display for Red and Yellow alarms.
• Red Only means that a fast alarm strip will display only for Red alarms.
Click a in the circle preceding the option to select it.
Sector Setup
Layout
Controls for formatting the Patient Sectors on the Main Screen. It has the following fields
• # Sectors - Set the number of Patient Sectors displayed on the device’s Main Screen -- 4, 6, 8,
12, or 16. This cannot exceed the # of Patients in the Purchased Options field in Device Setup
and Support Information. Click the up/down arrows to the right of the box to change the number
in the box.
• 1 Column/2 Columns - # of columns of Patient Sectors on the Main Screen. 1 Column is only
available for 8 patients or less. Click in the circle preceding the desired selection.
• Waves per Sector - Sets the number of monitoring waveforms that appear in Patient Sectors.
The number of waves can be 3 for 4 Patient Sectors, 2 for 6 - 12 Patient Sectors, 1 for 16 Patient
Sectors on a single display, and 2 for 16 Patient Sectors with Dual Resting Display enabled.
• Show Alarm Limits - select whether the Heart Rate Alarm Limits set in Alarm Management
and Setup will be displayed in Patient Sectors. Clicking a check in the box causes the
Heart Rate Alarm Limits to be displayed. No check means they will not be displayed.
• Dual resting display - Selects whether both displays of a dual display Information Center will
display Patient Sectors. A check means that Patient Sectors will display on both displays when a
Patient Window or All Controls are not active. No check means that Patient Sectors will display
on the primary display only. This field is not active if 1 is selected as the # of Displays.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue Series and is the answer not in the manual?

Philips IntelliVue Series Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue Series
CategoryServer
LanguageEnglish

Related product manuals