EasyManua.ls Logo

Philips IntelliVue Series - Equipment Capability; Information Center (Local Database); Printers

Philips IntelliVue Series
522 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Network Design
4-10 Site Planning and Preparation
Printers HP LaserJet printers can be network connected to the network for shared printing
capability.
A summary of these design capabilities and limitations of Patient Care Network
components is given in Table 4-1.
Information
Center (Local
Database)
This section includes independent Information Center designs only, that is Information
Centers that are not connected to a IntelliVue Clinical Network.
Equipment
Capability
Requirements must be matched to the following Philips Patient Care Network equipment
capabilities.
Information Center capability can:
display data from up to 16 patient monitors
annunciate 22 alarms
store up to 150 alarm records
store up to 96 hours of patient data for:
waves
ST segments
trends
events
•each System Communications Controller can manage data for up to 6 Information
Centers and from up to 24 patient monitors.
Patient monitors for both systems can be hardwired or telemetry.
Telemetry mainframes can support 8 channels of telemetry monitoring.
These design capabilities and limitations are summarized in Table 4-2, Philips Patient Care
Network Component Capabilities and Limitations.
Table 4-1. Release E.01 System Capabilities
Description of Limit
M3150 Local
Database
M3169 Small
Database
M3154 Central
Database
Maximum Number of Information Centers per system 1 3 8
Maximum Number of Clients per system 0 3 8
Maximum Number of Database Servers per system 0 1 10
Maximum Number of beds per Information Center 16 4 16
Maximum Number of monitored beds per system 16 48 128
Maximum Number of networked printers 2 4 8
Table 4-2. Philips Patient Care Network Component Capabilities and Limitations
Component Device Capabilities Limitations
Patient Monitors Hardwired monitor 1 per patient
Telemetry monitors 1 per patient
Telemetry Mainframe 8 monitors per Mainframe See Note below

Table of Contents