EasyManua.ls Logo

Philips IntelliVue Series - Time Synchronization

Philips IntelliVue Series
522 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Large Networks
Software Installation & Configuration
6-3
Server is responsible for maintaining the Patient List. There is no message on the
Information Centers connected to the non-master Database Server that the Master is
offline.
Time
Synchronization
All patient monitoring network components are time synchronized. Single synchronization is
supported from a master time source located either within the Clinical Network or provided
externally by the hospital. The external time source is added as a network device as part of the
Equipment Setup configuration window. The following table illustrates the supported
configurations and the time synchronization rules:
Table 6-1. M3150 Information Center (Local Database) without External Time Source
System Component
Can Date/Time be set on this
component?
Expected Behavior on Time Change
Information Center Yes Immediate Synchronization
Bedside No Synchronized within 5 minutes
Table 6-2. M3150 Information Center (Local Database) with External Time Source
System Component
Can Date/Time be set on this
component?
Expected Behavior on Time Change
Information Center No Synchronized 0-15 minutes
Bedside No Synchronized within 5 minutes after Information
Center synchronization
Table 6-3. Database Server System without External Time Source
System Component
Can Date/Time be set on this
component?
Expected Behavior on Time Change
Database Server Yes Immediate synchronization
Information Center No Synchronized within 1 minute
Information Center Client No Synchronized within 1 minute
Bedside No All bedsides synchronized within 5 minutes after
Information Center synchronization
Table 6-4. Database Server System with External Time Source
System Component
Can Date/Time be set on this
component?
Expected Behavior on Time Change
Database Server No Synchronized 0-15 minutes
Information Center No Synchronized within 1 minute after Database Server
Information Center Client No Synchronized within 1 minute after Database Server

Table of Contents

Related product manuals