Specifications
Hardware Description
2-51
Conversion factors for power dissipation are the following:
BTU/hr = Watts x 3.413
Kcal/hr = Watts x 0.8598
Conversion factor for air conditioning requirements for power dissipation is:
air conditioning power dissipation
1 ton = 8900 BTU/hr = 3024 Kcal/hr
Customer
Supplied
Displays
The following table gives recommended requirements for displays and video cables connected
to Information Centers and Clients.
Note For displays and video cables not supplied by Philips as part of an Information Center system:
• The support user is responsible for all aspects of their selection, purchase, installation,
repair, and disposal
• Philips cannot assure their system compatibility
• Philips cannot assure compliance with the ECG aspect ratio and 25 mm/s +
10% sweep
speed specification of Information Center software, which complies with the ANSI/
AAMI EC-13 Standard for Cardiac Monitors, Heart-Rate Meters, and Alarms.
UPS for Database Server
(1000 VA, 230V)
M3883-60001 160 - 286 n/a 50 - 60
Power Distribution Module M3166-60000 100, 120 -127,
200, 200 - 240
Yes 50 - 60
Table 2-6. Electrical Specifications of Database Server Components
Philips Component
Product
Part #
Input Voltage
(VAC)
Manual
Switching
Required?
Input
Frequency
(Hz)
Dissipated
Power (max)
(Watts)
Table 2-7. Recommended Display Specifications for Information Center Displays
Specification Value
Screen Resolution 1280 x 1024
Vertical Refresh Rate 60 Hz