Troubleshooting
Maintenance, Troubleshooting, and Repair
7-37
System Events To access System events that are designated Error and Warning, follow the steps for
Application Events on page 7-27, except check System in Step 2. The following table
gives entries that can appear as System events.
Table 7-8. System Events
Source Description (Event Code) Corrective Action
SDN Error: 513, SDN RAM error Replace SDN Interface Card
Error: 514, SDN signature RAM error Replace SDN Interface Card
Error: 515, SDN autopoll error Replace SDN Interface Card
Error: 516, SDN TB2 error Replace SDN Interface Card
Error: 517, SDN chip access not in dead time at start of dead
time. [(No SYNC TAP, (i.e., no SDN connection to IC)]
Check SDN cable connection to the Philips PC
Error: 519, Background not complete Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 769, Host not responding to interrupt from SDN card. Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 770, Write to host did not complete - DATA LOST. Reboot Philips system. If problem persists,
replace SDN Interface Card
Error 771, Read from host did not complete. Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 896, The card was told to read 0 bytes from the SDN
driver
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 897 The driver is trying to send too much data to the
card. The transfer did not complete.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 898, The amount of data read by the card from the driver
does not agree with the expected count. The transfer did not
complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 899, The transfer count indicating the amount of data
that the driver wants the card to read exceeds the amount that
can be read by the AMCC chip. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 900, The transfer count indicating the amount of data
that the driver wants the card to read is old. The transfer did
not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 901 The message header contained in the driver to card
data is not a valid header. Transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 902, The size of the recorder message from the driver to
the card is incorrect. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 903, The driver to card read appeared to be successful
but the transfer count to read did not agree with the amount
read. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 904, The recorder ID in the driver to card recorder
message is not a valid ID. Data not sent to the recorder.
Reboot Philips system. If problem persists,
replace SDN Interface Card