Troubleshooting
Maintenance, Troubleshooting, and Repair
7-13
Second NIC Card on Database
Server/Information Center is not
configured correctly.
Verify the Second NIC Card is configured
with the appropriate settings. See “Network
Card and TCP/IP Settings” on page 4-23
Hospital ID/Department ID
configuration settings do not
match those of the receiving
device
Verify the configurations settings in the
Purchased Options and Support Information
Page and the Equipment Setup (clinical unit
setup) window match. See Purchased
Options and Support Information and
Equipment Setup in Chapter 6.
Symptom Possible Cause Action