Troubleshooting
Maintenance, Troubleshooting, and Repair
7-25
Shutdown Shutdown includes the following:
• !Shutdown-Normal tool for the normal shut down of software for rebooting
• !Desktop Tools provides access to the Windows Main Menu.
Maintenance Maintenance includes the following:
• Restart IIS Admin Service stops and restarts the IIS service, which
supports the Information Center Web functionality. This does not cause the
Information Centers to operate in local mode.
• Restart DBServ stops and restarts the DbServ process. This does not cause
loss of patient data and does not cause Information Centers to operate in local
mode.
• Sychronize Large Network Server List used to synchronize the server
list on non-master Database Servers.
• !Reset Database Log resets the SQL server transaction log files to optimize
performance. This does cause Information Centers to operate in local mode.
On Database Servers only.
• !Upgrade NLS Catalogs used to automatically update the NLS catalogs for
new releases of patient monitors. On Database Servers only.
These tools, except Configuration tools, and their use in troubleshooting problems
are described in detail in the following sections. Configuration tools are described in
Chapter 6.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue Series and is the answer not in the manual?

Philips IntelliVue Series Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue Series
CategoryServer
LanguageEnglish

Related product manuals