Troubleshooting
Maintenance, Troubleshooting, and Repair
7-39
SDN Error: 930, Plugged devices table does not match number of
devices marked as plugged in
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 931, Invalid rack interface number or rack slot number Connect correct or new rack to the Philips
system
Error: 935, A device other than a Morgana recorder is plugged
in
Remove incorrect plug-in module
Error: 938, Can’t find a device status entry that should be there Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 1553, SDN adapter incompatible hardware and firmware Replace SDN Interface Card. If problem
persists, reinstall Philips software.
Error: 1554, SDN adapter missing hardware Replace SDN Interface Card.
Error: 1793, Adapter never responded to reset command within
time-out period. Adapter has been declared dead.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 1794, Firmware data file janus.afc is not accessible. Reinstall Philips software.
Error: 1795, Invalid firmware found on adapter -- Loading new
firmware from local firmware file
No action required, normal operation when the
Philips system is started.
Error: 1796, Unable to create the symbolic link for janus.afc. Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 1797, No SDN adapter interrupts have occurred within
time-out period - Scheduling an adapter reset.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 1798, Performing driver initiated adapter reset. No action required, normal operation when the
Philips system is started.
Error: 1799, The SDN adapter is running. No action required, normal operation when the
Philips system is started.
Error: 1800, A user requested an adapter reset via the
SdnMonitor utility.
No action required, normal operation when the
Philips system is started.
Error: 1801, The driver was not called back to perform an
adapter reset within the predetermined amount of time. The
driver and adapter have been made inactive.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 1802, Timeout while waiting for the OS kernel to call
the driver back to perform an adapter reset - retrying kernel
call-back request.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 1803, The previous error has repeated. To avoid filling
the event log with the same error, this error will not be logged
until a different error, or no error, is encountered.
No action required, normal operation when the
Philips system is started.
Error: 1804, The adapter’s boot strap code (DBOOT) has been
started. The driver will attempt to restart the SDN adapter.
No action required, normal operation when the
Philips system is started.
Table 7-8. System Events
Source Description (Event Code) Corrective Action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue Series and is the answer not in the manual?

Philips IntelliVue Series Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue Series
CategoryServer
LanguageEnglish

Related product manuals