Configuring the System
8-14 Start-up and Configuration
Table 8-4. M3170 - Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform Expected Results
What to Record
on Service
Record
V
isual For installation, perform visual inspection of shipping
cartons and contents.
No visible damage
Expected answer is
“Yes”. If so, Visual test
passed.
V:P or V:F
where P = Pass
F = Fail
Power On: Power on all devices.
Observe software boot up.
Observe that support display shows that the DataServer
Startup was complete.
Devices power up into
expected status on
support display
PO:P or PO:F
where P = Pass
F = Fail
Performance: Verify purchased options are enabled.
Verify normal system operation.
Do LAN based monitors record and print requests respond
to user inputs?
Expected answers are
“Yes”.
If so, Performance test
passed.
P:P or P:F
where P = Pass
F = Fail
If #C14 HL7 Export is installed, ping by host name the
target HL7 client device
Expected answer is
“Yes”. If so,
Performance test passed.
PHL7:P or PHL7:F
where P = Pass
F = Fail
S
afety Is the device connected to a 78581A/B (SCC) or telemetry
mainframe?
If answer is “No”, no
Safety test required.
If answer is “Yes”,
move to next step.
S:NA
where NA = Not
required
Is there a 2nd earth ground connected to the SCC or
telemetry mainframe
Expected answer is
“Yes”.
If so, Safety test passed.
S:P or S:F
where P = Pass
F = Fail
Confirm that 78581A/B or telemetry mainframe has
passed Safety Test Block (see 78581A/B or Telemetry
Mainframe Test and Inspection Matrix).
If answer is “Yes”,
Safety test passed.
If answer is “No”,
move to next step.
S:P or S:F
where P = Pass
F = Fail
Perform 78581A/B or Telemetry Mainframe Safety
Test Block
Expected answer is
“Yes”. If so, Safety test
passed
S:P or S:F
where P = Pass
F = Fail