Setting-Up Tests
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
3-26
3Move the trial lens as close to the patient’s eye as possible without
touching the lashes.
If you are running a SWAP (Blue-Yellow) test, the visor beneath the
for
ehead rest must be extended. You should also allow the patient to
adapt to the yellow bowl for about 3 minutes before testing. See
Chapter (9), "Short-Wavelength Automated Perimetry (SWAP)," for
further details.
4Review the patient’s position in the video eye monitor. The cross
(+) should be in the center of the pupil. Adjust as necessary.
When the patient has been instructed properly and positioned com-
fortably, you are ready to begin testing.