Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Care and Cleaning
15-7
Power Table Fuses
There are fuses on the HFA II-
i
power table located at the base of the lifting column. These fuses
control power to the table itself and the instrument (assuming the instrument is plugged into the
outlet under the table). Information on the replacement fuses and proper ratings are posted next to
the fuse location described below.
Power table fuses
control the power
to the table
Underside of power supply
Figure 15.2 HFA II-i Power Table With Mounted Printer Showing Location of Fuses
100-120V~ Table 220-240V~ Table
Fuse: T8A, 250V~ Fuse: T6.3A, 250V~
Note: Your actual fuse ratings may vary. Replace fuses wit
h the same rating as the original fuse that
was supplied with your table.
WARNING: Do not power the table using extension cords and do not use multiple
portable socket outlets.