Setting-Up Tests
3-25
Positioning the Patient
To facilitate patient positioning, the chin rest is divided into
two cups: one designated for right eye
testing; the other for left eye testing.
Position the patient’s chin here when
testing the right eye
Position the patient’s chin here when
testing the left eye
1Instruct the patient to place his or her chin on the appropriate side
of the chin rest, then assist with bringing the forehead against the
forehead rest.
Have the patient slide the chair in close to the HFA II-
i
. Adjust the
table height to be as high as necessary to keep the patient sitting
comfortably erect, rather than bent over or leaning forward.
If available, pull the slider handle out to release the slider. Slide the
HFA II-
i
toward the patient to allow improved posture for the test.
Release the handle to lock the slider in place.
2Align the patient’s eye on the video eye monitor so that the pupil is
centered in the target. Press the chin rest control in the direction you
want the patient’s eye to move in the video eye monitor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss Humphrey II-i series and is the answer not in the manual?

Zeiss Humphrey II-i series Specifications

General IconGeneral
BrandZeiss
ModelHumphrey II-i series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals