STATPAC Analysis & Printing
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
7-20
will get Single Field Analysis printouts instead of the desired Overview printout, as described in Step
4 in Table 7.2 above.
Note: For Change Analysis, Glaucoma Change Probability, and GPA, if you want to combine 24-2
and 30-2 results on one printout, STATPAC will analyze only the Central 24 degrees. If you want an
analysis of the Central 30 degrees, do not combine 24-2 and 30-2 results for these printouts. The
10-2 test cannot be used in conjunction with 24-2 or 30-2 tests, and is not available in the Change
Analysis, Glaucoma Change Probability, or Guided Progression Analysis format.
Printing from Recall Last Test
The HFA II-
i
holds in temporary memory the last right eye and left eye tested (they need not be for
the same patient) until the instrument is powered off, at which time the temporary memory is
cleared.
1. From the Main Menu screen, select RECALL LAST TEST.
2. Select the test eye (right or left) or CANC
EL.
3. Follow the instructions previously outlined for p
rinting current test results.
Printing from View Test
You can print any file you have retrieve
d through the View Test feature.
1. From the Main Menu screen, select the
FILE FUNCTIONS
icon.
2. Select VIEW TEST.
3. Designate the Source and Directory Order, then press PROCEED.
4. Select the test you want to retrieve.
5. Choose PROCEED to display the test results.
6. Follow the instructions previously outlined for p
rinting current test results.
Printing Delay
Whenever the File Directory Box is open (see example), print
ing will not occur.
If you have started a printing function, and a File Directory box is on the screen, finish your selection
or press the
MAIN MENU
icon in order to allow printing to continue.