Work List Networking Error Messages
Error Message Structure:
“Error occurred while retrieving the work list.
<Specific condition>”
D
o you want to Retry?
Table I.5 below lists error messages related to W
o
rk Lists and their possible resolution.
Table I.5 Work List Networking Error Messages
If you see While you are Check the following
Error occurred while retrieving
the work list.
Server responded with error:
<erro
r description>
Do you want to retry?
Retrieving the work list. Your work list server returned an error. See if the <error
de
scription> points to a possible resolution (For example this
may occur if you try to query the work list with Accession
number and the work list server does not support querying
work list with Accession number). Retry work list retrieval.
Refer to the work list server's manual for a possible re
solution
of the problem.
Error occurred while retrieving
the work list.
No items in the work list.
Do you want to retry?
Retrieving the work list. No patients were present in the work list that matched the
qu
ery. Retry by changing the work list query.
Error occurred while retrieving
the work list.
Invalid work list.
Do you want to retry?
Retrieving the work list. The work list retrieved contained an invalid
message. Retry
work list retrieval.
Error occurred while retrieving
the work list.
Work list not found.
Do you want to retry?
Retrieving the work list. This error can occur in Retrieve Only mode if the work list
server ha
s not yet made the work list.
Make sure the work list server is running and retry.
Error occurred while retrieving
the work list.
Too many response items.
Do you want to retry?
Retrieving the work list. Work list server returned large number of items in the work
list. Narr
ow your query and retry.
Error occurred while retrieving
the work list.
A network error occurred.
Do you want to retry?
Retrieving the work list. Make sure the HFA is connected to the network.
Make sure the Work list server is running.
•Retry.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss Humphrey II-i series and is the answer not in the manual?

Zeiss Humphrey II-i series Specifications

General IconGeneral
BrandZeiss
ModelHumphrey II-i series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals