Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Networking Reference
I-15
Setting Up Manual Archiving
Use these steps to set up manual archiving.
1 Select MAIN MENU>
SYSTEM SETUP
>COMMUNICATIONS SETUP>ARCHIVE/RETRIEVE SETUP
to open the Archive/Retrieve Setup screen.
.
Figure I.16 Communications Setup Screen
2 Open the Archive drop-down box and select MANUAL ONLY as the archiving frequency.
3 A Remind drop-down box will appear below the Archive drop-down box. Either leave the
drop-down box setting at NONE, or select WEEKLY (as illustrated below) or MONTHLY archiving
reminders.
Figure I.17 Selecting Weekly Archiving Reminders
If you select WEEKLY as the archive reminder frequency, upon starting up the HFA II-
i
, you will
be reminded to archive your data manually 7 days after your last archive, and you will continue
to be reminded every day thereafter until you do an archive. If you select MONTHLY as the