Guided Progression Analysis (GPA)
8-33
After making changes on the Exam Selection – GPA screen (see Add or Remove an Exam from
GPA and Change the Baseline Selection below), press
PROCEED and printing will begin. If you
have selected more than one eye, the Exam Selection – GP
A screen will appear for the other eye.
Repeat your selections for the remaining eye.
At any time, you can select the
SELECT DEFAULT button to return the exam selection to GPA’s default
selection.
Selection Mode Button
The selection mode of the Exam Selection – GPA scr
een is always
in either Follow-up or Baseline
Mode. Follow-up Mode allows you to add or remove exams for GPA analysis. Baseline Mode allows
you to select a different Baseline. Select the
SELECTION MODE button to toggle between Follow-up
and Baseline mode.
Add or Remove an Exam from GPA
By default, when you enter the Exam Selection – GPA scr
een, the Selection mode is Follow-up. To
add or remove exams from GPA, enter Follow-up mode and click on each exam that you want to
add or remove. Clicking an exam toggles it between selected and not-selected. A right arrow (>)
indicates the exam is selected for GPA. A minus sign (-) indicates the exam has been removed from
GPA, as shown in Figure 8.16. Exams with no symbol preceding the name ar
e not included because
they are not valid Follow-up exams for t
he currently selected Baseline.
Figure 8.16 Exam Selection – Follow-Up Selection Mode