Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Kinetic Testing
13-5
Setting Parameters
Kinetic testing always uses the central fixation target.
Gaze Tracking and Blind Spot Monitoring are
not available. The current stimulus value setting will display in standard Goldmann notation (see
Appendix (E), "Goldmann Conversion Tables," for stimulus conversion values) in the upper, middle
portion of the Kinetic Test screen.
The default settings for the kinetic stimulus are:
• Stimulus Color: White
• Stimulus Value: I 2 E (20 dB)
• Stimulus Speed: 4 degrees per second
Unless the default settings are changed, Kinetic testing will begin in
the peripheral field at
approximately 75° nasally and temporally, or 55° in the superior and inferior fields.
Each Kinetic test can have up to ten (10) sets of par
ameters that define different isopters or static
point stimuli. The results of the first isopter are marked on the Kinetic test field with the uppercase
letter “A”. The second isopter is indicated by a “B” on the display. Each subsequent new set of
parameters is then marked sequentially by the next letter of the alphabet.
1From the Kinetic Test screen, press CHANGE PARAMETERS.
2Use the drop-down menus to make any needed changes. See the
following pages for explanations.
Press SELECTION COMPLETE when done. You will return to the Kinetic
Test screen.