Kinetic Testing
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
13-34
5When the Static Points procedure is complete, press END STATIC
POINTS to return to the Special Mapping screen.
Custom Scan
The Humphrey Field Analyzer II-
i
allows you to move stimuli either from peripheral-to-central, from
central-to-peripheral, or from point-to-point within the field without going through the fixation
point. This feature may be helpful in describing the boundaries of a hemianopsia for example. Using
the Cursor method is recommended for selecting the Start and End points of the Custom Scan and
the method is described in the following pages. Selecting the Start and End points with the Keypad
is available but involves the entering of two coordinates (Meridian and Radius) for each point and,
therefore, takes more time.
1From the Special Mapping screen, select CUSTOM SCAN.
2You will see the Custom Scan screen. Select CHANGE PARAMETERS
if desired. Set the parameters. Review “Setting Parameters,” on
page 13-5 for details. Press SELECTION COMPLETE.