Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Kinetic Testing
13-35
3First, press the USE CURSOR button if the cursor is not available.
Locate the Start point by moving the Cursor by your preferred method
(finger on touch screen or with the glidepad) to the point from which
you wish to start the Custom Scan. Use the Meridian and Radius
locations at the top of the screen for greater accuracy.
4Press SELECT START POINT. A diamond appears at the position you
have chosen.
5Move the cursor to the area where you wish the test point to run
towards (the End Point).