Introduction & Instrument Setup
1-9
User Changes to Software or Hardware
The HFA II-
i
is a medical device. The software and hardware have been designed in accordance with
U.S., European and other international medical device standards designed to protect clinicians,
users and patients from potential harm caused by mechanical, diagnostic or therapeutic failures.
WARNING: Unauthorized modification of HF
A II-i software or hardware (including
peripherals) can jeopardize the safety of operators and patients, the performance of
the instrument, and the integrity of patient data; it also voids the instrument
warranty.
Approved Software
Use of software supplied or approved by Carl Zeiss Meditec for the HFA II-
i
is authorized. For the
current list of approved software call Carl Zeiss Meditec Customer Care: In the U.S., call
800-341-6968. Outside the U.S., contact your local Carl Zeiss Meditec distributor.
Note: Carl Zeiss Meditec does not provide technical support for the use of unappr
oved third party
software.
Instrument Installation
Only an authorized Carl Zeiss Meditec service representative should install the
HFA II-
i
.
In consultation
with the buyer, Carl Zeiss Meditec schedules one free on-site installation appointment after
instrument delivery. System installation and operator training require approximately one-half
business day.
Care in Handling
Use extreme care when handling and transporting the HFA II-
i
shipping boxes. The instrument
contains fragile optics that have been precisely aligned at the factory.
Installation Requirements
• The HFA II-
i
should operate on a dedicated power outlet. Based on your specification, we
configure your HFA II-
i
at the factory to use either 100V, 115V, or 230V line voltage.
• An isolation transformer is required when connecting peripher
al devices that are not Medical
Device approved (i.e., printer, USB drive, etc.) within 1.5 meters (4.9 feet) away from the
patient, such that the patient cannot touch a peripheral device with any part of his or her body
while being examined.
Tips to Avoid Damage
Note: Users are not authorized to dismantle or modify the HFA II-
i
hardware. To transport the
instrument outside the office, you must consult with a Carl Zeiss Meditec service technician. Failure
to do so voids all warranties offered with the HFA II-
i
.
• Only Carl Z
eiss Meditec authorized technicians should disassemble or service this instrument.
In the case of malfunction, error messages or operational problems, call Carl Zeiss Meditec
Customer Care: In the U.S., call 800-341-6968. Outside the U.S., contact your local Carl Zeiss
Meditec distributor.