EasyManua.ls Logo

Zeiss Humphrey II-i series - Importing Work Lists from DICOM Systems Using DICOM Gateway 2.0

Zeiss Humphrey II-i series
580 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Networking
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
14-48
Figure 14.54 Patient Data Screen with Patient Data and Accession Number
7 Select PROCEED to save the patient information and return to the Main Menu (or to a selected
test if you entered the Patient Data screen through a selected test).
Note: If a problem occurs during this procedure, refer to Ta
ble I.5, "Work List Networking Error
Messages," on page I-36.
Note: See “Imported Patient Rules and Conflicts,” on page 14-59 for information on imported
Patient ID and Issuer of ID rules and conflicts.
Importing Work Lists from DICOM Systems using DICOM
Gateway 2.0
Note: In order to use DICOM Gateway 2.0 work lists, WORK LIST must be selected in the DICOM
Gateway Services screen (see “Enable/Disable DICOM Gateway Services:,” on page H-4).
To import patient data using a work list
from your DICOM system using DICOM Gateway 2.0:
1 From the Main Menu, press the
PATIENT DATA
icon to display the Patient Data 1 screen as
shown in Figure 14.55.

Table of Contents

Related product manuals