Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
DICOM Gateway 2.0 (Optional)
H-3
3.
Configure your HFA to use the DICOM Gateway 2.0, enable services, and perform
connection tests to the DICOM servers (see “DICOM 2.0 Gateway Setup on HFA,” on
page H-3).
4.
Configure the HFA’s print destination (see “Setting Up Printing To a File,” on page I-25).
5.
Set up your Save and Transmit Options for the DICOM Archive to export a report at the
end of an exam (see “Setting Up Save & Transmit for EMR/PMS/DICOM Systems,” on
page I-24).
DICOM 2.0 Gateway Setup on HFA
To set up DICOM Gateway 2.0 o
n the HFA II-
i
:
1 Select Main Menu > System Setup > Communications Setup > DICOM Gateway Setup to
display the DICOM Gateway Setup screen (Figure H.1).
Figure H.1 DICOM Gateway Setup Screen
Set your local HFA settings:
2 Select INSTITUTION NAME.
3 Input a total of up to 64 characters and spaces, using the pop-up keyboard. Press ENTER.
4 Select STATION NAME.
5 Input a total of up to 16 characters and spaces, using the pop-up keyboard. Press ENTER.
6 Select AE TITLE.