Short-Wavelength Automated Perimetry (SWAP)
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
9-12
Case 2: (Continued)
Figure 9.7b Standard White-on-White Results for the SITA-SWAP Patient Shown in Figure 9.7a
This is the standard white-on-white visual field printout for this patient as discussed on the opposite
page. Here the GHT indicates the visual field is “Within Normal Limits” while the SITA-SWAP
indicates “Outside Normal Limits”. This patient was confirmed to have glaucomatous changes that
were recognized early with SITA-SWAP testing.