Reference to Older Test Strategies
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
L-2
STATPAC Test Parameters
STATPAC will analyze tests that fall within the parameters listed below:
Table L.1 STATPAC Parameters for White-on-White Perimetry
Type of test: Threshold
Test pattern: Central 10-2, 24-2, 30-2
Test strategy: SITA Standard, SITA Fast, Full Threshold
a
a. These strategies are still available on your HFA II-
i
. However, they have been replaced with SI-
TA-based testing in most practices.
, FastPac
a
Stimulus color: White
Stimulus size: Size III
Fixation target: Any
Foveal threshold: On or Off
Test speed: Normal or Slow
STATPAC analysis may be used with all Central 24-2 and 30-2
thr
eshold test results. There are some
limitations. For Central 10-2 test results, STATPAC produces a Single Field Analysis or an Overview
showing up to sixteen (16) tests results; the Change Analysis and Glaucoma Change Probability
Analysis are not available.
The parameters needed for STATPAC analysis of SWAP test re
sults are listed below. Single Field
Analysis and Overview printouts are available. The GHT is not available with FastPac tests.
Table L.2 STATPAC Parameters for Blue-Yellow Perimetry
Type of test: Threshold
Test pattern: Central 24-2, 30-2
Test strategy: SITA-SWAP (Central 24-2 pattern only), Full Threshold
a
a. These strategies are still available on your HFA II-
i
. However, they have been replaced with SITA-based
testing in most practices.
,
FastPac
a
Stimulus color: Blue
Stimulus size: Size V
Fixation target: Any
Foveal threshold: On or Off
Fluctuation test: On or Off (SITA-SWAP tests automatically are set to Off)
Test Speed: Normal or Slow