Short-Wavelength Automated Perimetry (SWAP)
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
9-14
Note: The “Abnormally High Sensitivity” alert will be seen on the SITA-SWAP printout if the false
positive rate exceeds 15%. However, it will not be triggered if only the overall height of the hill of
vision is elevated.
Licensing SITA-SWAP on Your HFA
The primary benefit to be gained from using SITA-SWAP testing is the ability to conduct SWAP
testing more quickly than would be possible when using standard SWAP testing. To add this time
advantage to your own HFA II-
i
, contact Carl Zeiss Meditec to purchase the required license for
SITA-SWAP software.
To obtain a SITA-SWAP license, you must contact Carl Zeiss Meditec:
In the U.S.: Call
Carl Zeiss Meditec at 1-800-341-6968.
Outside the U.S.: Con
tact your local Carl Zeiss Meditec distributor.
Once you have obtained your license, you can activate
it using the procedure that is provided in
Appendix (J) beginning with “Licensing GPA, SITA-SWAP, HFA-NET Pr
o, or DICOM Gateway 2.0,” on
page J-5.
Note: On an HFA II (Rev. 14.0 or higher system software) or HFA II-
i
(Rev. 4.0 or higher system
software), if SITA-SWAP is not licensed, you still can view/print SITA-SWAP exams that are generated
on a licensed HFA, but you cannot administer them.
Note: For offices with more than one HFA, if you import a SITA-SWAP exam to a recipient HFA that
carries a lower software revision (that does not acknowledge SITA-SWAP), attempted printouts of
SITA-SWAP data will be blank.