Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Kinetic Testing
13-17
The HFA II-
i
also will keep track of stimuli that were not seen. At the
end of the isopter, if any stimuli were unseen, this option will be
presented. Press YES if you wish to have these points retested. The
visual field will be reduced to the Central size to retest the missed
point. You will also be given the option of using the trial lens holder at
this point. You do not need to use it.
Pausing the Test
1You can pause the test at any time by pressing the STOP STIMULUS
button.
2Choose the action you wish to take while the test is paused. The
same meridian will be tested to completion when the test is resumed.
Note: There is no way for the patient to pause the test when
performing kinetic perimetry. Holding down the patient response
button will be recorded as a response to the stimulus. You must stop
the kinetic test manually, if you wish to give the patient a break.