Kinetic Testing
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
13-16
4Press AUTOMATIC.
You will then need to select the eye
to test. The Patient Data screen
appears. Press the PROCEED button when all the desired patient data
has been entered.
5Press START to begin the automatic test. The meridian to be tested
is indicated on the screen.
6The stimulus will light but will not move for a short period of time.
You then will see a small diamond begin to move across the screen
and the STOP STIMULUS button will become active.
When the patient presses the response button, the isopter letter will
appear and the next meridian to be tested will display on the screen.
The HFA II-
i
will pause briefly, then start to move the stimulus. This is
the routine throughout the entire test.