EasyManua.ls Logo

Zeiss Humphrey II-i series - HFA Network: Manual Setup (Static IP)

Zeiss Humphrey II-i series
580 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Networking Reference
I-7
3 You now need to enter your Workgroup or Domain name. You can do so by pressing the
WORKGROUP NAME button. Note: In order to use a Shared Folder, you must enter a
Workgroup Name.
4 Select OK to accept the name.
5 Select ALLOW REMOTE CONFIGURATION so that an X is displayed in the button if you want to
allow the
EasyConnect
RCT to configure the HFA II-
i
instrument.
6 Select HFA NAME to display a pop-up keyboard and enter (or change) the name (up to 24
characters) you want for this HFA II-
i
instrument. Press ENTER.
Note: It is strongly recommended to name your HFA II-
i
instrument for ease in identification when
using the
EasyConnect
RCT. Creating a name for an instrument can also allow you to know the
location of the instrument.
7 If an error message appears, refer to “Network Troubleshooting Error Messages,” on page I-31
for assistance.
Note: Any time you change any of the IP addresses on this scre
en, it is necessary to power down the
HFA II-
i
and then restart.
HFA Network: Manual Setup (Static IP)
If you wish to a manual network setup (static IP),
use the following procedure.
1 From the HFA Network Setup screen, select MANUAL from the OBTAIN HFA IP
CONFIGURATION drop-down box
(see Figure I.7).
Figure I.7 HFA Network Setup Screen – Manual
2 Select the HFA IP ADDRESS button and enter the IP address.
3 Select ENTER and the HFA’s IP address will appear to the right of the HFA IP ADDRESS button.

Table of Contents

Related product manuals