HFA II-i
Serial Cable
HFA I or HFA II
File Functions
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
10-20
Figure 10.8 Serial Communication Between an HFA I or HFA II and an HFA II-i
You can use a serial cable to transfer data between the three different series of HFA perimeters in
the following ways:
HFA I to an HFA II-
i
HFA II to an HFA II-
i
HFA II-
i
to and from an HFA II-
i
(Figure 10.9)
HFA II-i
HFA II-i
Serial Cable
Figure 10.9 Serial Communication Between an HFA II-i to and from an HFA II-i
CAUTION: HFA II-i series system software 5.x and the 15.0 system software release for
HFA II instruments create a different database structure than previous versions that is
not fully compatible with previous software releases.
WARNING: Every patient has an identifier that consists of Name, Date of Birth, Patient
ID, and Issuer of ID. To avoid unintentional automatic merging of patient records,
VERIFY a unique combination of Name, Date of Birth, Patient ID, and Issuer of ID
exists when patients are copied, transferred, retrieved from an archive, or restored
from a backup. There is no way to separate the combined records of erroneously
merged patients. Automatic merging of patient records occurs when a patient exists
in the local database with the EXACT SAME Name, Date of Birth, Patient ID, and
Issuer of ID as a patient that is copied, transferred, retrieved from an archive, or
restored from a backup.
For example, consider the following scenario:
Patient A: This patient is currently on the local HFA database and was automatically
assigned a unique Patient ID by the instrument because it was a blank field prior to a
software upgrade to V5.x software or during an import to the HFA through a copy,
transfer, retrieval from archive, or a restore from backup.