Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
EasyConnect RCT 1.0
G-3
Enable the RCT and Name HFA II-
i
Instruments
Each HFA II-
i
instrument must have “Allow Remote Configuration” selected to allow the RCT to
configure the instrument. This setting is selected by default. In addition, it is strongly recommended
to name your HFA II-
i
instruments for ease in identification when using the RCT. Creating a name for
an instrument can also allow you to know the location of the instrument.
To name your HFA II-
i
instruments and allow them to be configured with the RCT, follow the steps
below:
1 On the HFA II-
i
instrument, from the Main Menu, select
SYSTEM SETUP > COMMUNICATIONS
SETUP > HFA NETWORK SETUP
to display the HFA NETWORK SETUP SCREEN (Figure G.1).
Figure G.1 HFA Network Setup Screen
2 Select ALLOW REMOTE CONFIGURATION so that an X is displayed in the button. This allows the
RCT to configure the HFA II-
i
instrument.
3 Select HFA NAME to display a pop-up keyboard and enter (or change) the name you want for
this HFA II-
i
instrument. Press ENTER.
4 Select SAVE to save your changes.