EasyManuals Logo

Philips IntelliVue Series User Manual

Philips IntelliVue Series
522 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #316 background imageLoading...
Page #316 background image
Troubleshooting
7-30 Maintenance, Troubleshooting, and Repair
dataserver
redirector
recmgr
sdprocess
pdsstore
docstatusserver
Error: Unhandled
Exception Filter is
kicking in
Fatal error. System will automatically
reboot to clear error state
Reboot system to clear. Contact Response
Center if it happens often.
EMSMainMenu Winsock connection failed Reboot system. If this persists, reinstall OS
software.
Mode of operation could not be read from registry. Therefore,
the mode was defaulted to monitoring mode.
No action required.
The user logon flag could not be read from the registry.
Therefore, the user logon was defaulted to the
SystemManager account
Reboot system
Could not open the specified event to send to the Shell
application.
Reboot system
Could not start application because Error = XX Reboot system. If this persists, reinstall
Philips software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not write value to Registry because Error = XX Reboot system. If this persists, reinstall
Philips software
Install Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system.
Could not write value to Registry because Error = %d Reboot system. If this persists, contact
Response Center
OpenProcessToken failed. Contact Response Center
AdjustTokenPrivileges enable failed. Contact Response Center
Auto Reboot Failed Reboot system or recycle power
NetCfgLib Auto Reboot failed Reboot system, If this persists, reinstall
Philips software
OpenProcessToken failed Reboot system. If this persists, reinstall
Philips software
AdjustTokenPriveleges enable failed Reboot system. If this persists, reinstall
Philips software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not write value to Registry because Error = XX Reboot system. If this persists, reinstall
Philips software
Could not read value from Registry because Error = %d Reboot system. If this persists, reinstall
Philips software
Could not write to Error File No action required
OrbPlus Error: all No action required
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation Corrective Action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue Series and is the answer not in the manual?

Philips IntelliVue Series Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue Series
CategoryServer
LanguageEnglish

Related product manuals