Troubleshooting
Maintenance, Troubleshooting, and Repair
7-29
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation Corrective Action
dataserver Error: 2000 Monitoring conflict or Central Station
# conflict. A bed is being monitored by
more than one Information Center
Correct SDN branch assignments in SDN
Config and Display Setup to assure that
each monitor is assigned to only one
Central Station
Error: 2001 Patient Data Store told DataAcquisition
to tune more than 16 beds
simultaneously
Contact Response Center
Error: 2002 Network error. SCC probably
disconnected or turned off.
- Check SCC cable and SDN interface
connection
- Check that SCC is turned on.
Error: 2004 Length of message block received from
Redirector does not equal the first word
of the buffer
Contact Response Center
Error: 2005 Invalid link to next SDN message.
Corrupt incoming data.
Contact Response Center
Error: 2006 SDN message too small Contact Response Center
Error: 2007 HRC check failed. Corrupt incoming
data
Contact Response Center
Error: 2008 Invalid link to next SDN message.
Corrupt incoming data.
Contact Response Center
Error: 2009 Invalid high order bits in SDN
message. Corrupt incoming data.
Contact Response Center
Error: 200B Virtual Terminal Protocol could not
connect to arrhythmia controls object.
Contact Response Center
Error: 200C Virtual Terminal Protocol could not
connect to database alarms object
Contact Response Center
Error: 200D Virtual Terminal Protocol error in
GetAlarmWave
Contact Response Center
Error: 200E, 200F,
2010, 2011, 2106,
2017, 2018, 2019
Low level Data Control Protocol error No action required
Error: 201A Data Acquisition Receive thread is
shutting down
Not an error, no action required
Error: 201B Low level Data Control Protocol error No action required
Error: 2021 Branch number changed. User altered
SCC configuration while system was
running
- Check SCC cabling
- Shut down Information Center and restart.
System should recognize new
branch and start normally.
Error: 2022 SDN Interface Card reset Shut down Information Center and restart
Error: 2023 Data Link Control header incorrectly
terminated
Reseat or replace SDN Interface Card
Error: 2024 SDN Data block incorrectly terminated Reseat or replace SDN Interface Card

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue Series and is the answer not in the manual?

Philips IntelliVue Series Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue Series
CategoryServer
LanguageEnglish

Related product manuals