Acquire Spectroscopy Data
53
Figure 6.2. Spectroscopy - Alternate Scan Direction: Disabled
Figure 6.3. Spectroscopy - Alternate Scan Direction: Enabled
2. Spectra Acquisition specifications.
This group of settings deals with the number of spectra and the times and delays within each spectrum
determination.
The first edit box allows you to specify the Number of Spectra to Acquire.
The next parameter, the Pre-sample Delay, is the interval you want to wait before each Spectral
response signal acquisition begins
The final Spectra Acquisition parameter is the Sampling Time. This value together with the data
channel sampling rate determines the number of Spectral Response Signal measurements that are
averaged together to determine the recorded value of the Spectral Response Signal for the current
level of the independent parameter. (R9 also holds the entire time series separately in the History file,
this feature unique to R9 provides you with extra information about the spectral response signals often
critical to interpretation of unusual spectra.)