EasyManua.ls Logo

RHK Technology R9 - Figure 6.9. Spectroscopy Settings and Circuit Diagram for the Trial DI;DV Spectroscopy; Figure 6.10. Modification to Pre-Configured UHV300 Dashboard

Default Icon
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquire Spectroscopy Data
58
connected and the Gate acting as the diode cathode. A series resistor of 100 Megohms was used to set
the current level appropriate for using the IVP-200 as the input preamplifier.
Figure 6.9. Spectroscopy Settings and Circuit Diagram for the Trial dI/dV
Spectroscopy
In your case, you will have a real tunneling probe in operation. Modify the above settings to reflect the
correct values for your system. Then complete an I/V trace at a single location. It should resemble this
above result. Repeat the Bias Spectroscopy experiment until you are confident in these settings. Use
these settings as the basis for the dI/dV experiment.
This experiment is implemented in the following way.
1. Modify the Lock-In component of the Dashboard as shown below.
Figure 6.10. Modification to Pre-configured UHV300 Dashboard
The Lock-in configured for STM should resemble the diagram. The marked settings control the
modulation and demodulation behavior.
2. Choose the modulation frequency on the Bias Modulation panel.

Table of Contents