STRASBAUGH, INC.
6-14 February 19, 2002 7AF SECS/GEM Interface Ver. 1.9
Strasbaugh Confidential
The Host can delete all report definitions and associated links in the Equipment by using the following special format:
S2F33 W * H->E
<L [2]
<U4 DATAID> * DATAID
<L>
> .
6.34 S2F34 - Define Report Acknowledge
S2F34 * H<-E
<B [1] DRACK> . * DRACK - Acknowledge Code
Normally, DRACK is zero. Any non-zero DRACK indicates the Equipment has rejected the entire S2F33 message.
DRACK Description
0x00 OK
0x02 Denied. Invalid format.
0x03 Denied. At least one RPTID already defined.
0x04 Denied. At least one VID does not exist.
6.35 S2F35 - Link Event Report
S2F35 W * H->E, Multiblock
<L [2]
<U4 DATAID> * DATAID
<L
<L [2]
<U4 CEID> * Collection Event ID
<L
<U4 RPTID> * Report ID
...
>
>
...
>
> .
The Host links Report Ids to selected Collection Event Ids. The Specified CEIDs are initialized to "disabled". See S2F37
for enabling CEIDs.
If S2F35 is multi-block, the Host may optionally send the S2F39/S2F40 Inquire/Grant transaction before sending S2F35,
but the 7AF does not require it.
The Equipment ignores DATAID.
The Host can eliminate all links for a CEID by sending the following format in S2F35 for that CEID:
...
<L [2]
<U4 CEID> * Collection Event ID
<L>
>
...