Short-Wavelength Automated Perimetry (SWAP)
9-7
4. As with white-on-white perimetry, we recommend that you confirm any abnormal findings
before making a diagnosis that is based on the visual field results. Patients often exhibit a
“learning effect” and may be better prepared for the SITA-SWAP testing routine on a second
visit. This leads to an improved recognition of the blue stimulus, as well as better pacing of
patient responses. These factors combine to provide more accurate test results.
5. It is possible to use the Full Threshold and FastP
ac testing strategies to do SWAP. However, we
do not recommend that you use these strategies, as the test times can be two or three times
longer than those of a SITA-SWAP test. Further information on Full Threshold and FastPac test
strategies is presented in Appendix (L).