EasyManua.ls Logo

Analog Devices ADuCM356 - Page 158

Analog Devices ADuCM356
312 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual ADuCM356
USE CASE CONFIGURATIONS
analog.com Rev. A | 158 of 312
Figure 43. R
CAL
Measurement
The R
CAL
measurement is used to calibrate the R
SENSOR
+ R
LOAD
measurement in Step 3 by using ratiometric measurements. As
shown in Figure 43, the electrochemical sensor is biased via the
low-power potentiostat and low-power TIA amplifiers for Step 4.
The waveform generator and the high-speed DAC generate a 10
mV amplitude sine waveform.
The AC excitation loop D node, P node, and RCAL0 pin are shorted
by setting the Dx switch and Px switch to close the DR0 and PR0
switches (SWCON, Bits[7:0] = 0x11). Other Dx switches and Px
switches are open. The AC excitation loop N node, high-speed TIA
T node, and RCAL1 pin are shorted by setting the Nx switches
and Tx switches to close the NR1 and TR1 switches (SWCON,
Bits[15:8] = 0x85). Close T9 to select R
TIA02
(SWCON Bit 17 =
1). Other Nx switches and Tx switches are open. The stimulus is
added to the 200 Ω R
CAL
and the high-speed TIA. The ADC signal
chain and DFT block are used to calculate the impedance of R
CAL
.
Set the low-power loop switches to allow the excitation amplifier
and the high-speed TIA to set the DC bias voltage level of the
sensor. The following example code demonstrates how to configure
the switches in the low-power loop:
AfeLpTiaSwitchCfg (channel,
SWMODE_AC); //Low Power Loop for AC impe►
dance switch settings (0x3180)

Table of Contents