Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
EasyConnect RCT 1.0
G-29
10After all instruments have been configured, the RCT will display a Commit screen (Figure G.33)
before sending configuration messages to any instrument
. This screen gives you a last chance
to make corrections before committing the specified changes. Select
BACK to go back to the
previous screen(s) to make any corrections. Selecting
CANCEL will exit the RCT without
configuring any of the selected instruments.
Figure G.33 Commit Screen
Select the PERFORM CONNECTION TESTS checkbox to perform a connection test for each config-
ured setup. Results of these tests will be presen
ted on the Completion screen and a VIEW TEST
LOG
button will be provided if any instrument failed a connection test.
When you select
NEXT, all selected instruments will be sent the specified configuration.