Networking Reference
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
I-34
Could not update local TCP/IP settings. Verify IP Address, Subnet Mask, and Gateway Address (Note:
Most likely cause is Gateway Address is not within the Subnet
Mask).
Error occurred accessing the r
emote folder. Press OK to
continue. Press Cancel to return to the setup menu.
Press OK and retry to gain access. Change folders if necessary, or
get needed access rights from the System Administrator.
Press CANCEL to return to the previous menu.
Error occurred accessing the remote folder
. Press OK to save
the settings. Press Cancel to return to the setup menu.
Press OK to save the settings to the hard drive.
Press CANCEL to end the process and
return to the setup menu.
Error occurred while retrieving the work list. Retry the operation.
Failed to ping the remote host. Verify that the remote host is powered and active. Then, retry the
oper
ation.
Failed to replace database with File Server backup. Free up hard drive storage space on the HFA II-
i
, as necessary.
Retry loading the backup to the hard drive.
Not enough hard disk space to backup to file server. Make
mor
e space by deleting old exams.
Free up HFA II-
i
hard drive storage space, as necessary.
Not enough hard disk space to perform archive. Make more
space by delet
ing old exams.
Free up hard drive storage space on the HFA II-
i
, as necessary.
Not enough hard disk space to perform
data retrieval. Make
more space by deleting old exams.
Delete old exams from HFA II-
i
hard drive and retry retrieving data
from the server.
The database backup was empty. A restore was not
performed
.
You have not performed your first database backup to the server.
The database has not yet been ar
chived. You can archive data
by selecting Archive/Retrieve from the File Functions menu.
You have not performed your first database archiving to the
server.
Archive data by selecting ARCHIVE/RETRIEVE fr
om the File
Functions menu.
The database is empty. A backup was not created. There is no data available for backing up on the hard drive of the
HF
A II-
i
.
The Internet gateway IP address should be within the
specified subnet.
Please correct the gateway IP address.
Enter the correct gateway IP address.
Unknown error occurred, archive not completed. Repeat the process. If it fails on second and subsequen
t attempts,
write down the details of your process and then contact Carl Zeiss
Meditec Customer Service.
Unknown error occurred, backup to file server not completed. Repeat the process. If it fails on second and subsequen
t attempts,
write down the details of your process and then contact Carl Zeiss
Meditec Customer Service.
Table I.4 General Networking Error Messages
Error Message Possible Solution