0: Common commands Model DMM7510 7½ Digit Graphical Sampling Multimeter
6 DMM7510-901-01 Rev. B / May 2015
You cannot combine the command sets.
Example
*LANG?
Set the command set to TSP.
Verify setting by sending the command set query.
Output:
TSP
The TSP command set is in use.
Also see
Status model (on page 1)
*OPC
This command sets the operation complete (OPC) bit after all pending commands, including overlapped
commands, have been executed.
Type Affected by Where saved Default value
Usage
*OPC
*OPC?
Details
When *OPC is sent, the OPC bit (bit 0) in the Status Event Status Register is set after all pending
command operations have been executed. After all programmed operations are complete, the
instrument returns to idle, at which time all pending commands (including *OPC and *OPC?) are
executed. After the last pending command is executed, the OPC bit is set or an ASCII “1” is placed in
the Output Queue.
When the trigger model is executing, most sent commands are not executed. If a command cannot
be processed, an error event message is generated in the event log.
Also see
:INITiate[:IMMediate] (on page 6-177)
opc() (on page 8-230)