Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8: TSP command refer
ence
DMM7510-901-01 Rev. B / May 2015 8-323
trigger.model.setblock() — trigger.BLOCK_NOTIFY
This function defines a trigger model block that generates a trigger event and immediately continues to the next
block.
Type TSP-Link accessible Affected by Where saved Default value
Function Yes
Restore configuration
Instrument reset
Power cycle
Configuration script Not applicable
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_NOTIFY, trigger.EVENT_NOTIFYN)
blockNumber
The sequence of the block in the trigger model
N
The identification number of the notification; 1 to 8
Details
When trigger model execution reaches a notify block, the instrument generates a trigger event and
immediately continues to the next block.
Other commands can reference the event that the notify block generates. This assigns a stimulus
somewhere else in the system. For example, you can use the notify event as the stimulus of a
hardware trigger line, such as a digital I/O line.
Example
digio.line[3].mode = digio.MODE_TRIGGER_OUT
trigger.model.setblock(5, trigger.BLOCK_NOTIFY, trigger.EVENT_NOTIFY2)
trigger.digout[3].stimulus = trigger.EVENT_NOTIFY2
Define trigger model block 5 to be the notify 2 event. Assign the notify 2 event to be the stimulus for digital
output line 3.
Also see
Notify block (on page 3-83)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB