7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-43
DISPlay subsystem
This subsystem contains commands that control the front-panel display.
:DISPlay:CLEar
This command clears the text from the front-panel USER swipe screen.
Type Affected by Where saved Default value
Usage
:DISPlay:CLEar
Example
DISP:SCR SWIPE_USER
DISP:USER1:TEXT "Batch A122"
DISP:USER2:TEXT "Test running"
Clear the USER swipe screen and switch to
display the USER swipe screen.
Set the first line to read "Batch A122" and
the second line to display "Test running".
Also see
:DISPlay:USER<n>:TEXT[:DATA] (on page 6-47)
:DISPlay:<function>:DIGits
This command determines the number of digits that are displayed for measurements on the front panel.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
See
Usage
:DISPlay:<function>:DIGits <n>
:DISPlay:<function>:DIGits DEFault
:DISPlay:<function>:DIGits MINimum
:DISPlay:<function>:DIGits MAXimum
:DISPlay:<function>:DIGits?
:DISPlay:<function>:DIGits? DEFault
:DISPlay:<function>:DIGits? MINimum
:DISPlay:<function>:DIGits? MAXimum
The function to which the setting applies; see Functions
3.5 digit resolution:
4.5 digit resolution: 4
5.5 digit resolution: 5
6.5 digit resolution: 6
7.5 digit resolution:
(not available for digitize functions)