8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-10 DMM7510-901-01 Rev. B / May 2015
acal.lastrun.time
This attribute returns the date and time when auto calibration was last run.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
dateTime = acal.lastrun.time
dateTime
The date and time
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
• MM/DD/YYYY is the month, date, and year
• HH:MM:SS.NNNNNNNNN is the hour, minute, second, and fractional second
Example
print(acal.lastrun.time)
Returns the date and time when the auto
calibration was last run.
Example output:
05/28/2014 00:37:47.743000000
Also see
acal.run() (on page 8-12)
Auto calibration (on page 3-44)