7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-63
[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh
This command defines the threshold value for the pulse width.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
20 µs
Usage
[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh <setting>
[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh?
<function>
The function to which the setting applies; see Functions
<setting>
The threshold value for the pulse width: 1 µs to 40 ms
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This option is only available when the analog trigger mode is set to pulse.
This option sets either the minimum or maximum pulse width that generates an analog trigger event.
The value of pulse condition determines whether this value is interpreted as the minimum or
maximum pulse width.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE PULSE
CURR:ATR:PULS:LEV 2.5
CURR:ATR:PULS:POL BEL
CURR:ATR:PULS:WIDT 30e-6
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to pulse.
Set the analog trigger level to 2.5 A.
Set the polarity to below.
Set the analog trigger pulse width to 30
µ
s.
Also see
Analog triggering overview (on page 3-64)
[:SENSe[1]]:<function>:ATRigger:MODE (on page 6-59)
[:SENSe[1]]:<function>:ATRigger:PULSe:CONDition (on page 6-60)
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel (on page 6-61)