8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-238 DMM7510-901-01 Rev. B / May 2015
scriptVar.save()
This function saves the script to nonvolatile memory or to a USB flash drive.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
scriptVar.save()
scriptVar.save(filename)
The name of variable that references the script
The file name to use when saving the script to a USB flash drive
Details
The scriptVar.save() function saves a script to nonvolatile memory or a USB flash drive. The
root folder of the USB flash drive has the absolute path /usb1/.
If no filename is specified, the script is saved to internal nonvolatile memory. If a filename is
given, the script is saved to the USB flash drive.
If you set scriptVar to autoexec, the script is run when the instrument powers up. You must
delete the existing autoexec script before saving the new one. Note that performing a system reset
does not delete the autoexec script.
If no filename is specified (the filename parameter is an empty string), the script is saved to internal
nonvolatile memory. If a filename is given, the script is saved to the USB flash drive.
You can add the file extension, but it is not required. The only allowed extension is .tsp (see
Example 2).
Example 1
Saves the script referenced by the variable
to nonvolatile memory.
Example 2
test8.save("/usb1/myScript.tsp")
Saves the script referenced by the variable
test8 to a file named myScript.tsp on your
Also see
Working with scripts (on page 7-5)