Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-281
trigger.lanout[N].logic
This attribute sets the logic on which the trigger event detector and the output trigger generator operate on the
given trigger line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
trigger.LOGIC_NEGATIVE
Usage
logicType = trigger.lanout[N].logic
trigger.lanout[N].logic = logicType
logicType
The type of logic:
• Positive: trigger.LOGIC_POSITIVE
• Negative:
trigger.LOGIC_NEGATIVE
N
The LAN event number (1 to 8)
Example
trigger.lanout[2].logic = trigger.LOGIC_POSITIVE
Set the logic for LAN trigger line 2 to
positive.
Also see
None
trigger.lanout[N].protocol
This attribute sets the LAN protocol to use for sending trigger messages.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
lan.PROTOCOL_TCP
Usage
protocol = trigger.lanout[N].protocol
trigger.lanout[N].protocol = protocol
protocol
The protocol to use for messages from the trigger:
• lan.PROTOCOL_TCP
• lan.PROTOCOL_UDP
•
lan.PROTOCOL_MULTICAST
N
The LAN event number (1 to 8)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals