6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-22 DMM7510-901-01 Rev. B / May 2015
Example
ACAL:SCH?
Sets auto calibration to run every day at 8 am
and query to verify the settings.
Output:
Also see
:ACAL:RUN (on page 6-20)
CALCulate subsystem
The commands in this subsystem configure and control the math and limit operations.
:CALCulate2:<function>:LIMit<Y>:AUDible
This command determines if the instrument beeper sounds when a limit test passes or fails, or disables the
beeper.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Other functions: NONE
Usage
:CALCulate2:<function>:LIMit<Y>:AUDible <state>
:CALCulate2:<function>:LIMit<Y>:AUDible?
The function to which the setting applies; see Functions
• Never: NONE
• On test failure: FAIL
• On test pass:
Functions
Details
The tone and length of beeper cannot be adjusted.