7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-347
tsplink.initialize()
This function initializes all instruments and enclosures in the TSP-Link system.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
nodesFound = tsplink.initialize()
tsplink.initialize()
tsplink.initialize(expectedNodes)
nodesFound
The number of nodes actually found on the system, including the node on which
the command is running
expectedNodes
The number of nodes expected on the system (1 to 32)
Details
This function regenerates the system configuration information regarding the nodes connected to the
TSP-Link system. You must initialize the system after making configuration changes. You need to
initialize the system after you:
• Turn off power or reboot any instrument in the system
• Change node numbers on any instrument in the system
• Rearrange or disconnect the TSP-Link cable connections between instruments
If the only node on the TSP-Link network is the one running the command and expectedNodes is
not provided, this function generates an error event. If you set expectedNodes to 1, the node is
initialized.
If you include expectedNodes, if nodesFound is less than expectedNodes, an error event is
generated.
Example
nodesFound = tsplink.initialize(2)
print("Nodes found = " .. nodesFound)
Perform a TSP-Link initialization and indicate how many
nodes are found.
Example output if two nodes are found:
Nodes found = 2
Example output if fewer nodes are found and if
localnode.showevents = 7:
1219, TSP-Link found fewer nodes than
expected
Nodes found = 1
Also see
localnode.showevents (on page 8-227)
tsplink.node (on page 8-351)
tsplink.state (on page 8-352)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals