8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-282 DMM7510-901-01 Rev. B / May 2015
Details
The LAN trigger listens for trigger messages on all the supported protocols. However, it uses the
designated protocol for sending outgoing messages.
After you change this setting, you must re-connect the LAN trigger event generator before you can
send outgoing event messages.
When multicast is selected, the trigger IP address is ignored and event messages are sent to the
multicast address 224.0.23.159.
Example
print(trigger.lanout[1].protocol)
Get LAN protocol that is being used for sending
trigger messages for LAN event 1.
Also see
trigger.lanout[N].connect() (on page 8-278)
trigger.lanout[N].ipaddress (on page 8-280)
trigger.lanout[N].stimulus
This attribute specifies events that cause this trigger to assert.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
trigger.EVENT_NONE
Usage
LANevent = trigger.lanout[N].stimulus
trigger.lanout[N].stimulus = LANevent
LANevent
The LAN event that causes this trigger to assert; see Details for values
N
A number specifying the trigger packet over the LAN for which to set or query
the trigger source (1 to 8)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals