Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-240 DMM7510-901-01 Rev. B / May 2015
:TRIGger:TIMer<n>:STARt:SEConds
This command configures an alarm or a time in the future when the timer will start.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings 0
Usage
:TRIGger:TIMer<n>:STARt:SEConds <time>
:TRIGger:TIMer<n>:STARt:SEConds?
<n>
Trigger timer number (1 to 4)
<time>
The time: 0 to 2,147,483,647 s
Details
This command configures the alarm of the timer.
When the timer is enabled, the timer starts immediately if the timer is configured for a start time that
has passed.
Example
TRIG:TIM1:STAR:SEC 60
TRIG:TIM1:START:FRAC 0.5
TRIG:TIM1:STAT ON
Set the timer for 60.5 s.
Enable the trigger timer for timer 1.
Also see
:TRIGger:TIMer<n>:STATe (on page 6-242)